Network Isolators – Answers to frequently asked questions:

Baaske Medical network isulators meet the high requirements of EN 60601 and can be used in medical technology and medical facilities such as hospitals or doctors' practices in the patient environment.

Find out everything you need to know about network isulators here in the FAQ.

Not found the right answer? We will be happy to help you by e-mail at sales@e-medic.com or by phone at 05741 236027-70.

Network Isolators – Answers to frequently asked questions: Baaske Medical network isulators meet the high requirements of EN 60601 and can be used in medical technology and medical... read more »
Close window
Network isolators

Netzwerkisolator_MI1005_15a7c1b2eec78b

Network Isolators – Answers to frequently asked questions:

Baaske Medical network isulators meet the high requirements of EN 60601 and can be used in medical technology and medical facilities such as hospitals or doctors' practices in the patient environment.

Find out everything you need to know about network isulators here in the FAQ.

Not found the right answer? We will be happy to help you by e-mail at sales@e-medic.com or by phone at 05741 236027-70.

Network isolators are used for the galvanic isolation of a copper-guided Ethernet data line. Via electromagnetic induction, high-frequency alternating voltages in the frequency band used for data transmission are transmitted with as little loss as possible. Due to this process, the passive, electronic components do not require their own power source and behave completely transparently in the data network.
A network isolator in the medical field / medical network disconnector serves to protect the patient from leakage currents. The network connection between an active medical device and an Ethernet network or a non-medical device, such as an office PC or laptop, is safely galvanically isolated in accordance with the IEC EN 60601-1 standard (3rd edition) for medical electrical devices and systems.
Why a single network insulator cannot be a medical device and what an operator should know.
What does a network insulator do and why does it have to be used?
If a medical device is connected to a non-ME device, e.g. a printer, the created data line must be galvanically isolated in accordance with the requirements of IEC 60601-1 or EN 60601-1 and a network isolator must be used.
Installation errors, ageing processes, environmental influences or moisture can make cable sheaths brittle or fragile. The unwanted electrical connection to other live parts can lead to high overvoltages and voltage peaks for a short time and lead to a fault within the electrical system.
Network isulators must be used wherever electrically conductive connections via network cables are not permitted. This can be in commercial as well as in public and private areas. The usual areas of application are:
Here you can find out for which user groups the MED MI1005 network isolators have been developed:
Worth knowing for manufacturers of medical devices with network interface.
CE marking of network isolators. Devices for galvanic isolation of Ethernet data signals, so-called network isolators, are used as part of a medical electrical system consisting of active medical products (e.g. ultrasound or endoscope) and not medical products (e.g. computers, switches).
Application example network isolator MI 1005 using on a gastroscopy cart. How to use a network isolator and why is the use of an isolating transformer is insufficient.
EN 60601-1-2:2015 - Teil 1-2: Elektromagnetische Verträglichkeit medizinischer elektrischer Geräte. Begriffserklärung.